ONH836系列:氧氮氢分析仪(熔融法)
软件特点和优势
Cornerstone氧氮氢分析软件分成4个主要部分:分析、诊断、设定、仪器,界面可以快速导航和组织。工具条、滑动条和下拉式菜单简化了校正和数据处理的参数设定。分析软件同时包括实时的环境参数的监测,提供完整的系统图表模拟。高级诊断程序包括全数字化的在线操作手册、维护模拟演示、图片说明等提供快速指导。Cornerstone氧氮氢分析软件同时支持***的数据文档格式和可变的报告格式。 可选具有压力补偿和温度控制性能的标气检测气路。山东美国LECO分析仪多少钱
Procedure – Powder/Chip Samples
Determine the instrument blank.
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode either manually or remove the crucible and press the analyze button to clean with an automatic cleaner if applicable.
Add approximately 0.05 g of *** Graphite Powder to a 782-720 Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Instrument calibration/drift correction.
a. Login a minimum of 3 Standard replicates.
b. Weigh a ~1.000 gram of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
c. Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
d. Place the calibration/drift standard into the open port at the top of the loading head.
e. Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Procedure - Solid Samples
Instrument calibration/drift correction.
Place the calibration/drift standard in a *** Nickel Basket, and if applicable, place the sample into the appropriate autoloader position.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
Note: samples using automation should be placed in the appropriate autoloader position before starting the
analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
Procedure - Solid Samples
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically
提供各种无机材料、钢铁、有色金属、硬质合金、陶瓷材料中宽范围的氧氮氢同时分析。山东美国LECO分析仪多少钱
仪器的亮点和优势:
特点和优势
·炉头区域LED照明系统
·炉头快速拆装的粉尘过滤装置
·装有自动旁路阀的可快速拆装氧水捕集装置
·**的动态流量补偿(DFC)保证热导池气体流量、压力恒定
高效脉冲炉冷却系统
·强力内循环水冷装置热交换效率
·双直流风扇保证持续冷却,不受电压波动的影响
·***电极炉头设计,提高脉冲炉
全自动分析系统(可选)
·自动清扫装置
·20位或100位全自动分析系统
·高效全自动真空除尘装置直接将粉尘排至仪器外部捕集器中,消除环境污染
改进的红外检测池技术
·***红外池的恒温绝热技术消除了外界环境温度的干扰
·长寿命的红外发射光源和无漂移的电路设计保证红外检测池的长期稳定性
·氧高低双量程采用3个碳红外检测器同时检测一氧化碳和二氧化碳,保证准确定量 山东美国LECO分析仪多少钱
Procedure–Powder/ChipSamplesInstrumentcalibration/...
【详情】Procedure-SolidSamplesInstrumentcalibration/driftc
【详情】Procedure–Powder/ChipSamplesDeterminetheinstrument...
【详情】氧氮氢分析仪的常见故障及解决方法供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度...