Procedure - Solid Samples
Determine the instrument blank.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of 3 Standard replicates.
Weigh approximately *** grams of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
Note: LECO Reference Materia***o not require preparation.See preparation statement on the reference material certificate.
新型气路管道减少气路连接阀,使得漏气,断路故障更少。山东美国LECO分析仪咨询报价
Instrument Highlights and Features
Features and Benefits
LED light ring illuminates the furnace area
Choice of either argon or helium carrier gas
Novel electrode design increases heat transfer efficiency for increased stability
Non-touch interface systems are also available
User-friendly brand software
High-Performance Detector Design
Thermostatic construction provides increased protection from ambient temperature fluctuations
Optimized emitter control and detection circuitryimproves the IR cell lifetime and long-term stability,
resulting in superior accuracy and precision
Infrared detection for the determination of oxygen, thermoconductivity detection for the determination
of nitrogen
山东美国分析仪咨询报价稳固的炉头试样加载结构。
ONH836系列:氧氮氢分析仪(熔融法)
软件特点和优势
Cornerstone氧氮氢分析软件分成4个主要部分:分析、诊断、设定、仪器,界面可以快速导航和组织。工具条、滑动条和下拉式菜单简化了校正和数据处理的参数设定。分析软件同时包括实时的环境参数的监测,提供完整的系统图表模拟。高级诊断程序包括全数字化的在线操作手册、维护模拟演示、图片说明等提供快速指导。Cornerstone氧氮氢分析软件同时支持***的数据文档格式和可变的报告格式。
Procedure - Solid Samples
Prepare the instrument as outlined in the operator's instruction manual.
Determine the instrument blank.
Login a minimum of 3 Blank replicates.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Place a 502-344 Nickel Basket into the loading head.
Note: samples using automation should be placed in the appropriate autoloader position before starting the
analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
User-Friendly Brand Software Cornerstone
LECO's exclusive brand software enables the user to have complete access to analysis Cornerstone control, method settings, diagnostics, reporting, and more in a highly organized and immersive environment. Designed through a collaboration of customer feedback and innovative engineering,Cornerstone allows the user to conduct all of their day-to-day operations within a single Analysis screen designed for speed and ease-of-use. Our innovative grouping of sample data into sets and replicates simplifies the data output and automatically calculates relevant statistics, alleviating the need for additional data processing.
减少了繁琐的额外数据处理。新疆专业分析仪
力可创新的批次和重复样品数据分组功能极大简化了样品结果输出和结果标准偏差的自动计算。山东美国LECO分析仪咨询报价
Procedure – Powder/Chip Samples
Determine the instrument blank.
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode either manually or remove the crucible and press the analyze button to clean with an automatic cleaner if applicable.
Add approximately 0.05 g of *** Graphite Powder to a 782-720 Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.