Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a***Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen,the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3i a minimum of three times for each calibration/drift standard used.
长寿命的红外发射光源和无漂移的电路设计保证红外检测池的长期稳定性。江苏元素分析仪厂家直销User-Friendly Brand Software Cornerstone
LECO's exclusive brand software enables the user to have complete access to analysis Cornerstone control, method settings, diagnostics, reporting, and more in a highly organized and immersive environment. Designed through a collaboration of customer feedback and innovative engineering,Cornerstone allows the user to conduct all of their day-to-day operations within a single Analysis screen designed for speed and ease-of-use. Our innovative grouping of sample data into sets and replicates simplifies the data output and automatically calculates relevant statistics, alleviating the need for additional data processing.
湖南氧氮氢分析仪厂家供应氧氮氢分析仪选哪家?给你推荐上海禹重科技!
Oxygen, Nitrogen, and Hydrogen in Refractory Metals
Summary
One of the most critical chemical specifications of titanium alloys is the hydrogen content. Too high of a
hydrogen content can cause hydrides to precipitate, which can lead to embrittlement and subsequent
cracking when the alloy is stressed. Hydrogen pickup typically occurs during downstream processing steps
such as heat treating, pickling, and cleaning.
The LECO ONH836 is a simultaneous oxygen, nitrogen, and hydrogen determinator that utilizes an electrode furnace, inert carrier gas, and both infrared and thermal conductivity detection to meet the analytical needs of the refractory metal industry.
This application note was written specifically for use with the LECO ONH836 series determinator.
Procedure - Solid Samples
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically
Procedure - Solid Samples
Prepare the instrument as outlined in the operator's instruction manual.
Determine the instrument blank.
Login a minimum of 3 Blank replicates.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Place a 502-344 Nickel Basket into the loading head.
Note: samples using automation should be placed in the appropriate autoloader position before starting the
analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
长寿命的红外发射光源和无漂移的电路设计保证检测器的长期稳定性。江苏元素分析仪厂家直销
Procedure - Solid Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.
Calibrate/drift following the procedure outlined in the operator's instruction manual.