Instrument Highlights and Features
Features and Benefits
LED light ring illuminates the furnace area
Choice of either argon or helium carrier gas
Novel electrode design increases heat transfer efficiency for increased stability
Non-touch interface systems are also available
User-friendly brand software
High-Performance Detector Design
Thermostatic construction provides increased protection from ambient temperature fluctuations
Optimized emitter control and detection circuitryimproves the IR cell lifetime and long-term stability,
resulting in superior accuracy and precision
Infrared detection for the determination of oxygen, thermoconductivity detection for the determination
of nitrogen
全自动分析系统(可选) 。广东专业分析仪咨询报价
Instrument calibration/drift correction
f. Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.
g. Firmly place a graphite crucible on the lower electrode tip.
h. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
i. Repeat steps 3b through 3h a minimum of three times for each calibration/drift standard used.
. Calibrate/drift following the procedure outlined in the operator's instruction manual.
氧氮氢分析仪的常见故障及解决方法
供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度太高或仪器通道电压不正常。如没有水流,炉子温度太高这些信息会显示在显示器画面上,但没有信息显示说明这两项正常。接下来检查仪器通道零位电压,如果比±3V高出1V以上,可能是因为气瓶空了,或者是空气进入到分析仪中。检查并更换化学试剂,如果有空气进入热导池里,热导池的电压就会<-6V,此时打开右面的门,堵住炉子气体进口,10s后,热导池电压值必然升高。经过逐一排查,**终确认碱石棉有问题,更换后仪器正常。
Procedure – Powder/Chip Samples
Prepare the instrument as outlined in the operator's instruction manual.
Login a minimum of three Blank reps.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Insert a *** Nickel Capsule (leave capsule open) into the open port at top of loading head.
Note: samples using automation should be placed in the appropriate autoloader position before starting the analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
Procedure - Solid Samples
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically
提供各种无机材料、钢铁、有色金属、硬质合金陶瓷材料中宽范围的氧氮氢同时测定。广东专业分析仪咨询报价
Sample Preparation
Typically, titanium and zirconium samples are chemically etched to remove surface contamination when oxygen and nitrogen are determined. However, etching can introduce hydrogen into the sample.
ASTM method E 1409 "Determination of Oxygen and Nitrogen in Titanium and Titanium Alloys by the Inert
Gas Fusion Technique", as updated in 1996, permits either etching or abrading (filing) of the test specimen.
ASTM E 1447 "Determination of Hydrogen in Titanium and Titanium Alloys by the Inert Gas Fusion Thermal
Conductivity/Infrared Detection Method" permits surface preparation by abrading (if necessary to remove
contamination). Differences in sample preparation present somewhat of a dilemma regarding simultaneous
determination of O, N, and H in titanium. However, abrading samples with a file to remove surface
contamination will yield accurate O, N, and H results. The ONH836 utilizes a high-power electrode furnace to quickly and efficiently release the target gases from within the sample, which allows for a very rapid simultaneous determination of oxygen, nitrogen, and hydrogen.
广东专业分析仪咨询报价上海禹重实业有限公司致力于仪器仪表,是一家服务型的公司。公司业务涵盖光谱仪,色谱仪,质谱仪,样品前处理等,价格合理,品质有保证。公司秉持诚信为本的经营理念,在仪器仪表深耕多年,以技术为先导,以自主产品为重点,发挥人才优势,打造仪器仪表良好品牌。禹重科技立足于全国市场,依托强大的研发实力,融合前沿的技术理念,飞快响应客户的变化需求。
Procedure–Powder/ChipSamplesInstrumentcalibration/...
【详情】Procedure-SolidSamplesInstrumentcalibration/driftc
【详情】Procedure–Powder/ChipSamplesDeterminetheinstrument...
【详情】氧氮氢分析仪的常见故障及解决方法供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度...
【详情】Instrumentcalibration/driftcorrectionf.Cleantheupp
【详情】Procedure–Powder/ChipSamplesDeterminetheinstrument...
【详情】SamplePreparationTypically,titaniumandzirconiumsam
【详情】InstrumentHighlightsandFeaturesOptionalFeaturesOpt
【详情】